Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 31: (1 year) CAL VER /5700,5725 INSTRUMENT: Fluke 33: (1 year) CAL VER /5700,5725 DATE: 02-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 26 NUMBER OF LINES: 207 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= # # Source: # Fluke 31/33 Instruction Manual (PN 896118 Sep. 1991). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 5725_3x.bmp # 57_3x.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F W 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP This procedure requires a 50-turn Toroid Coil. 1.004 DISP Refer to the 31/33 Instruction Manual figure 7, page 32 1.004 DISP for directions for constructing a suitable coil. 1.005 HEAD {RMS AMPS TESTS} 1.006 DISP Turn the UUT on. 1.007 PIC 5725_3x 1.008 HEAD RMS AMPS TESTS: {400A Range}, 400A @ 60Hz 1.009 TOL 400P% 0.0085U 1.010 5700 8.00A 60H B1 S 2W 1.011 ACC 400.0A TOL 1.012 MEMI Enter UUT reading in amps. 1.013 MEME 1.014 MEMC 400 A 8.2U 60H 2.001 HEAD RMS AMPS TESTS: 400A Range, 200A @ 60Hz 2.002 5700 4.00A 60H B1 S 2W 2.003 ACC 200.0A TOL 2.004 MEMI Enter UUT reading in amps. 2.005 MEME 2.006 MEMC 400 A 4.2U 60H 3.001 JMPT 3.005 Fluke 33: (1 year) CAL VER /5700,5725 3.002 DISP Disconnect the toroid from the 5725A. 3.003 PIC 57_3x 3.004 JMP 14.004 3.005 HEAD {PEAK AMPS TESTS} 3.006 DISP Press the SMOOTH/CREST button twice to select PEAK. 3.007 HEAD PEAK AMPS TESTS: {400A Range}, 400A @ 60Hz 3.008 TOL 400P% 0.012U 3.009 5700 8.00A 60H B1 S 2W 3.010 ACC 565.8Ap TOL 3.011 MEMI Enter UUT reading in amps. 3.012 MEME 3.013 MEMC 400 Ap 17.2U 60H 4.001 HEAD PEAK AMPS TESTS: 400A Range, 200A @ 60Hz 4.002 5700 4.00A 60H B1 S 2W 4.003 ACC 282.85Ap TOL 4.004 MEMI Enter UUT reading in amps. 4.005 MEME 4.006 MEMC 400 Ap 8.75U 60H 5.001 DISP Disconnect the toroid from the 5725A. 5.002 PIC 57_3x 5.003 HEAD PEAK AMPS TESTS: 400A Range, 100A @ 60Hz 5.004 TOL 650P% 0.0028U 5.005 5700 2.00A 60H S 2W 5.006 ACC 141.45Ap TOL 5.007 MEMI Enter UUT reading in amps peak. 5.008 MEME 5.009 MEMC 400 Ap 4.55U 60H 6.001 HEAD PEAK AMPS TESTS: 400A Range, 80A @ 50Hz 6.002 5700 1.60A 50H S 2W 6.003 ACC 113.1Ap TOL 6.004 MEMI Enter UUT reading in amps peak. 6.005 MEME 6.006 MEMC 400 Ap 3.7U 50H 7.001 HEAD PEAK AMPS TESTS: {40A Range}, 40A @ 50Hz 7.002 5700 0.80A 50H S 2W 7.003 ACC 56.6Ap TOL 7.004 MEMI Enter UUT reading in amps peak. 7.005 MEME 7.006 MEMC 40 Ap 2.0U 50H 8.001 HEAD PEAK AMPS TESTS: 40A Range, 20A @ 400Hz 8.002 5700 0.40A 400H S 2W 8.003 ACC 28.285Ap TOL 8.004 MEMI Enter UUT reading in amps peak. 8.005 MEME 8.006 MEMC 40 Ap 4.55U 400H 9.001 HEAD PEAK AMPS TESTS: 40A Range, 10A @ 400Hz 9.002 TOL 700P% 3.54e-4U 9.003 5700 0.20A 400H S 2W 9.004 ACC 14.145Ap TOL 9.005 MEMI Enter UUT reading in amps peak. 9.006 MEME 9.007 MEMC 40 Ap 0.865U 400H 10.001 HEAD PEAK AMPS TESTS: 40A Range, 8A @ 60Hz 10.002 TOL 700P% 3.54e-4U 10.003 5700 0.16A 60H S 2W 10.004 ACC 11.31Ap TOL 10.005 MEMI Enter UUT reading in amps peak. 10.006 MEME 10.007 MEMC 40 Ap 0.64U 60H 11.001 HEAD PEAK AMPS TESTS: 40A Range, 4A @ 60Hz 11.002 5700 0.08A 60H S 2W 11.003 ACC 5.66Ap TOL 11.004 MEMI Enter UUT reading in amps peak. 11.005 MEME 11.006 MEMC 40 Ap 0.47U 60H 12.001 HEAD PEAK AMPS TESTS: 40A Range, 2A @ 50Hz 12.002 5700 0.04A 50H S 2W 12.003 ACC 2.825Ap TOL 12.004 MEMI Enter UUT reading in amps peak. 12.005 MEME 12.006 MEMC 40 Ap 0.385U 50H 13.001 HEAD PEAK AMPS TESTS: 40A Range, 1A @ 50Hz 13.002 TOL 700P% 3.54e-5U 13.003 5700 0.02A 50H S 2W 13.004 ACC 1.415Ap TOL 13.005 MEMI Enter UUT reading in amps peak. 13.006 MEME 13.007 MEMC 40 Ap 0.345U 50H 14.001 HEAD {RMS AMPS TESTS} 14.002 DISP Press the SMOOTH/CREST button to select RMS. 14.003 RSLT =400A Range 14.004 HEAD RMS AMPS TESTS: 400A Range, 100A @ 60Hz 14.005 TOL 650P% 0.002U 14.006 5700 2.00A 60H S 2W 14.007 ACC 100.0A TOL 14.008 MEMI Enter UUT reading in amps. 14.009 MEME 14.010 MEMC 400 A 2.2U 60H 15.001 HEAD RMS AMPS TESTS: 400A Range, 80A @ 50Hz 15.002 5700 1.60A 50H S 2W 15.003 ACC 80.0A TOL 15.004 MEMI Enter UUT reading in amps. 15.005 MEME 15.006 MEMC 400 A 1.8U 50H 16.001 HEAD RMS AMPS TESTS: {40A Range}, 40A @ 50Hz 16.002 5700 0.80A 50H S 2W 16.003 ACC 40.0A TOL 16.004 MEMI Enter UUT reading in amps. 16.005 MEME 16.006 MEMC 40 A 1.0U 50H 17.001 HEAD RMS AMPS TESTS: 40A Range, 20A @ 400Hz 17.002 5700 0.40A 400H S 2W 17.003 ACC 20.00A TOL 17.004 MEMI Enter UUT reading in amps. 17.005 MEME 17.006 MEMC 40 A 1.10U 400H 18.001 HEAD RMS AMPS TESTS: 40A Range, 10A @ 400Hz 18.002 TOL 700P% 2.5e-4U 18.003 5700 0.20A 400H S 2W 18.004 ACC 10.00A TOL 18.005 MEMI Enter UUT reading in amps. 18.006 MEME 18.007 MEMC 40 A 0.70U 400H 19.001 HEAD RMS AMPS TESTS: 40A Range, 8A @ 60Hz 19.002 5700 0.16A 60H S 2W 19.003 ACC 8.00A TOL 19.004 MEMI Enter UUT reading in amps. 19.005 MEME 19.006 MEMC 40 A 0.36U 60H 20.001 HEAD RMS AMPS TESTS: 40A Range, 4A @ 60Hz 20.002 5700 0.08A 60H S 2W 20.003 ACC 4.00A TOL 20.004 MEMI Enter UUT reading in amps. 20.005 MEME 20.006 MEMC 40 A 0.24U 60H 21.001 HEAD RMS AMPS TESTS: 40A Range, 2A @ 50Hz 21.002 5700 0.04A 50H S 2W 21.003 ACC 2.00A TOL 21.004 MEMI Enter UUT reading in amps. 21.005 MEME 21.006 MEMC 40 A 0.24U 50H 22.001 HEAD RMS AMPS TESTS: 40A Range, 1A @ 50Hz 22.002 TOL 700P% 2.5e-5U 22.003 5700 0.02A 50H S 2W 22.004 ACC 1.00A TOL 22.005 MEMI Enter UUT reading in amps. 22.006 MEME 22.007 MEMC 40 A 0.22U 50H 23.001 HEAD {FREQUENCY TESTS} 23.002 DISP Press the Hz button. 23.003 HEAD FREQUENCY TESTS: 40A Range, {10A @ 400Hz} 23.004 5700 400.0H 1.1U 0.20A 2W 24.001 HEAD FREQUENCY TESTS: 40A Range, {4A @ 60Hz} 24.002 5700 60H 0.4U 0.08A 2W 25.001 HEAD FREQUENCY TESTS: 40A Range, {1A @ 50Hz} 25.002 5700 50H 0.4U 0.02A 2W 26.001 END